D08-1FD Single Channel Display
The D08-1FD flow display instrument uses a small plastic housing, making it compact and easy to install. In addition to the standard 0~+5V voltage signal, the flow input and output signals also include 4~20mA or 0~10mA current signals. The D08-1F flow display instrument has flow setting and valve control functions and can be used with MFCs (or MFMs) with a working power supply of ±15V.
Product Description
Features
• 2.4-inch LCD screen, displays input/output signal type, real-time flow rate, flow rate setpoint, flow
• rate unit, valve control, cleaning, shutdown, and communication type indication;
• Flow rate can be set via a digital rotary encoder, allowing for step and continuous flow rate settings;
• The device automatically saves settings, preventing data loss during power outages;
• Uses silicone rubber buttons for aesthetics, comfort, and longevity
• Reserved RS485/RS232 interface, supports Modbus RTU protocol;
• Parameters can be configured or changed via host computer software, offering convenience and supporting calibration;
• Signal interface and usage are fully compatible with the original D08-1F;
• Uses a dedicated national standard AC connector for enhanced safety;
• Safety and EMC comply with relevant CE standards;
• Complies with RoHS environmental standards
Technical Parameters
Type | D08-1FD |
Input Power | ~85-265 VAC, 50/60Hz |
Output Power | +15V±3% 400mA, -15V±3% 400mA |
Rated Power | 15 W |
Input/Output Power Signal | Voltage signal: 0~+5V |
Current signal: 4~20mA or 0~10mA | |
Note: 0~+5V is the basic configuration; optional additional functions include 4~20mA or 0~10mA input/output | |
Buttons | Cleaning, Shutdown, Valve Control, Setting, Left Shift, Right Shift, Rotary Encoder |
Display Type | 2.4-inch LCD |
Settable display units: %, SCCM, SLM | |
Accuracy | 0.1 grade (0~5V, 4~20mA, 0~10mA) |
Temperature Coefficient | 0~50℃ AD, DA change ≤0.1% F.S. |
Operating Temperature Range | 0~50℃ |
Dimensions | 96×96×160 mm (panel opening 91.5*91.5mm) |
Communication Method | RS485/RS232, selectable via internal jumper cap |
Compatible with Modbus communication protocol, see protocol description for details | |
Control Channel | 1-channel MFC |
EMI Resistance | Complies with CE, RoHS certification standards |
Weight | ≤0.5 Kg |